| Technical data (20°C)  | 
        
            | Measurement range  | From 0… 100 mbar to 0… 40 bar vacuum, com- | Zero thermal drift ±0.025% of F.S / °C max.  | 
        
            |   | pound, gauge or absolute pressure  | Option: ±0.015% of F.S / °C max.  | 
        
            | Output signal  | E612: 0...10 Vdc  | Span thermal drift Typically: ±0.01%/°C / max.: ±0.015%/°C  | 
        
            |   | E613: 4...20 mA  | Wetted parts Ceramic + 1.4404 (316L) stainless steel + NBR  | 
        
            |   | E614: 1...5 Vdc  | seal (standard)  | 
        
            | Supply voltage  | E612: 14...40 Vdc  | Standard connections Electrical: Connector DIN 43650  | 
        
            |   | E613-E614: 8...40 Vdc  | Pressure: G1/2  | 
        
            |   | Protection against polarity inversion  | Many options available  | 
        
            | Insulation  | > 100 MΩ at 250 Vdc. Option: 500 Vdc.  | Protection rating (EN 60529) Standard: IP65 (Connector DIN)  | 
        
            | Maximum input current  | E612-E614: 6 mA  | Option: IP67 (depending on connector)  | 
        
            | Load impedance (+M/-M)  | E612: ≥ 2.5 k Ω  | Typical response time < 30 ms  | 
        
            |   | E614: ≥ 1 k Ω E613: (Usupply-24)/0,02 ≤ RΩ ≤ (Usupply-8)/0,02  | Vibration resistance (IEC 68-2-6) 1.5 mm (10 - 55 Hz), 20 g (55 Hz - 2 kHz) Shock resistance (IEC 68-2-32) 25 falls from 1 m on to a concrete floor  | 
        
            | CE Conformity  | 89/336 EMC directive CE with screened cable,  |   | 
        
            |   | screen connected at both ends  |   | 
        
            |   | 97/23/CE PED pressure directive  |   | 
        
            | Accuracy  | ±0.5% of F.S.  |   | 
        
            | Linearity, hysteresis, repeatability  | ±0.2% of F.S. (reference to BFSL)  |   | 
        
            | Operating temperature Ambient (Ta)  | -25...+85°C (only  |   | 
        
            | Option  | Low t°: -40...+85°C (only for seal NBR)  |   | 
        
            |   | High t°: -25...+100°C (only for DIN 43650 connector)  |   | 
        
            | Fluid temperature  | -25...+100°C (Ta ≤ 50°C)  |   | 
        
            | Storage temperature  | -40...+85°C  |   | 
        
            | Compensated temperature range -25...+85°C  | 
        
            | (zero and sensitivity)  |